Odyssey M Specifications
Imaging System Specifications
Unless otherwise specified, the following apply to Odyssey M Model 3350 and Odyssey M Model 3340.
Laser Source
The Odyssey M is a Class 1 laser product.
Laser Source | Maximum Laser Output | Beam Divergence |
488 nm | 30 mW | 0.038° |
520 nm | 50 mW | 0.046° |
685 nm | 450 mW | 0.085° |
785 nm | 1.2 W | 0.060° |
Strong Electromagnetic Fields
Operating the Odyssey M in the presence of a strong electromagnetic field in the 100MHz to 400MHz frequency range has been shown to degrade the performance of the chemiluminescent imager. It is possible that some allowable faint bars or striping may be seen at slightly more than the background levels of the camera. When used in the expected environment, this interference should be uncommon.
LED
LED Risk Group: Exempt group in accordance with IEC 62471:2006. The LEDs do not pose any photobiological hazard.
Focus
Focus offset range: -1.00 mm to 5.00 mm
Pixel Resolution
Available imaging resolutions: 5 μm, 10 μm, 20 μm, 50 μm, 100 μm
Resolution for chemi images: 100 μm
Maximum Height of Samples
To image an item on the Odyssey M, the item must be 23 mm or less from the Scan Surface to the top of the item. Tall samples could be pushed off the scan surface during imaging and fall into sensitive areas of the instrument that should not come into contact with samples. Special care should be taken with ex vivo samples.
Operating Specifications
Operating Conditions
For indoor use only; operating temperature
and dew point not greater than , non-condensing; altitude not to exceed 2000 m.Transport Conditions
Shipping/transport conditions for the Odyssey M Imager should not exceed -40 °C to 65 °C.
Environmental Conditions
Pollution Degree 2
Power Requirements
- Universal input range is between 100-240 VAC (voltage fluctuations not to exceed 10% of the nominal voltage); < 3 Amp maximum; At 120 V: 0.5 Amp typical when idle; 50-60 Hz.
- Power Supply: One end of the provided power supply connects to a wall outlet and the other end connects to the imager. The power supply converts from wall outlet voltage to 24 VDC (5 Amp). Only use the provided power supply to plug in the imager. If you need another power supply, contact LICORbio.
Dimensions
24" width × 30" depth × 15" height (61 cm width × 76 cm depth × 38 cm height)
Height with lid fully open: 28" (71 cm)
Scan Surface
Total Scan Area: 25 cm W × D 18 cm (9.8” W × 7.1” D)
Scan Area of Chemi Region (Odyssey M Model 3350 only): 15 cm W x 11 D cm (5.9” W × 4.3” D)
Dynamic Range
>6 logs for chemiluminescence (Model 3350) and fluorescence. This dynamic range is obtained in a single acquisition.
Weight
Odyssey M Model 3350: 121 lbs (55 kg)
Odyssey M Model 3340: 115 lbs (52 kg)
Network
- Protocol: TCP/IP
- Connection: Use the Category 6 network cable provided with the Odyssey M.
Detectors
sCMOS image sensor
Sensor for chemiluminescence (Model 3350): CCD
Expected Laser Lifetime
685 nm: 20,000 hours
785 nm: 20,000 hours
488 nm: 40,000 hours
520 nm: 40,000 hours
Light Sources
RGB LED (trans-illumination)
RGB LED (reflective illumination)
Solid-state diode laser at 488 nm
Solid-state diode laser at 520 nm
Solid-state diode laser at 685 nm
Solid-state diode laser at 785 nm
Odyssey M Detection Channel Summary
Channel | Excitation / Illumination | Emission Filter | Odyssey M Channel | Assay Category / Imaging Mode |
---|---|---|---|---|
800 | 785 nm | 816 - 840 nm | 785Ex-820Em | Membrane, In-Cell Western (includes On-Cell Western), Cell Analysis, , Custom |
700 | 685 nm | 721 - 740 nm | 685Ex-720Em | Membrane, Gel, In-Cell Western (includes On-Cell Western), Cell Analysis, , Custom |
700A | 685 nm | 816 – 840 nm | 685Ex-820Em | Custom |
520 | 520 nm | 570 - 610 nm | 520Ex-590Em | Membrane, Gel, In-Cell Western (includes On-Cell Western), Cell Analysis, , Custom |
520A | 520 nm | 721 - 740 nm | 520Ex-720Em | Custom |
520B | 520 nm | 816 - 840 nm | 520Ex-820Em | Custom |
488 | 488 nm | 519 - 543 nm | 488Ex-530Em | Membrane, Gel, In-Cell Western, Cell Analysis, , Custom |
488A | 488 nm | 570 - 610 nm | 488Ex-590Em | Gel, Custom |
488B | 488 nm | 721 - 740 nm | 488Ex-720Em | Custom |
488C | 488 nm | 816 - 840 nm | 488Ex-820Em | Custom |
Chemi | NA | NA | Chemi | Membrane, Custom (luminescent assays) |
RGB Epi | 470 525 630 Epi | NA | RGBEpi | Custom |
630 Epi | 631 - 659 nm | Clear Window | 630Epi | Membrane |
525 Epi | 500 - 530 nm | Clear Window | 525Epi | Membrane |
470 Epi | 447 - 472 nm | Clear Window | 470Epi | Custom |
RGB Trans | 470 525 630 Trans | NA | RGBTrans | Custom | ,
630 Trans | 631 - 659 nm | Clear Window | 630Trans | Gel, Absorbance Assay |
525 Trans | 500 - 530 nm | Clear Window | 525Trans | Custom |
470 Trans | 447 - 472 nm | Clear Window | 470Trans | Gel, Absorbance Assay |