Replacing the Chamber Shaft Seal

Printable PDF: Replacing the Chamber Shaft Seal

(8250_Replacing the Chamber Shaft Seal_21612.pdf)

Download this content as a pdf that can be saved to your computer or printed.

The shaft seal assembly on 8250-104/C Long-Term Chambers should be replaced if it fails. Failure is indicated by the mechanism squealing or shuddering during motion.

Components

The chamber shaft seal replacement kit (part number 8200-928) includes the following components.

Description Part Number
Shaft Seal Assembly 9982-025
Thread Lock 222 Purple (0.5 mL) 208-10201
Synthetic Oil (2 mL) 210-11015
Rheolube 368A (1 mL) 210-11016
  1. With the chamber powered on, press the OPEN/CLOSE button twice in succession to move the chamber to its “Park” position.
  2. Disconnect the chamber power/data cable.
  3. Use a 3/8" wrench to remove the arm mounting nuts on the underside of the arm.

    Collect the nuts, lock washers, and flat washers. Leave the cable and tubes connected and set the bowl assembly aside.

  4. Use a 9/64" hex key to remove the two hex bolts on the chamber adapter.

  5. Use a Phillips #1 screwdriver to remove the four screws and washers from the shaft seal assembly.

  6. Remove the shaft seal assembly from the drive shaft.

    If it is stuck, grasp it firmly and twist it loose. Remove the O-ring if it sticks to the housing.

  7. Clean any residue off of the top of the housing.

    Use a mild detergent, if needed. Do not use sharp objects, and do not damage the finished surface. Make sure the sealing surfaces are clean and smooth before installing the new shaft seal assembly.

  8. Lubricate the new shaft seal assembly with oil and grease from the shaft seal assembly kit.
    • Apply the full quantity of synthetic oil to the felt washer inside the shaft seal assembly.
    • Apply a generous layer of Rheolube to the O‑ring and rubber seal with your finger or a cotton-tipped swab.
  9. Be sure the O-ring is in place on the bottom of the shaft seal assembly.

  10. Install the new shaft seal assembly on the shaft.

    You may need to press firmly and twist the assembly to get the O-ring over the shaft, but use caution and be sure not to cut, pinch, or gouge the O-ring or seal.

  11. Apply a droplet of Loctite® from the spares kit to each of the four screws, then install the screws and washers to secure the assembly.

    Tighten the screws.

  12. Wipe any oil or grease off of the top of the shaft and then install the chamber adapter.

    The slotted side goes toward the chamber opening. Tighten the hex screws.

  13. Install the bowl and arm assembly using the nuts and washers.
    1. Tighten the nuts only finger tight, and then adjust the chamber bowl so that it is centered over the soil collar gasket.

    2. Tighten the two arm mounting.

You have completed the chamber shaft seal replacement. You can verify that the chamber operates correctly by connecting it to the power supply and opening and closing the chamber.